Model 709AL Positive Displacement Autosampler
The Alcott Model 709AL
Positive Displacement Autosampler is an automatic sampling and fixed volume
injection system. The 709AL requires no accessory gases or syringes and, while
using Alcotts' patented FilterCap
TM, will
automatically filter your sample "In-The-Vial". The 709AL can be equipped with
or without an injection valve to allow for optimum configuration for existing or
newly designed systems.
OPERATION
The benefits of the Positive Displacement technique are
many, including viscosity independence in sample transfer from vial to valve's
loop, the highest injection precision possible and simplicity of operation. The
technique, as demonstrated below, is such that a uniquely designed cap is
pressed into an Alcott Certified
TM 8mm vial which
contains sample. Sample is then transferred to the injection valve by the piston
action of the hollow autosampler needle which first penetrates the vial cap,
then uses the cap to displace the vial's contents.
BENEFITS
- 128 position sample vial tray
- Integrated high speed injection valve
- Selectable needle depth (4 positions)
- "On-the-fly" programming and editing of methods
- System interface through RS-232, IEEE, Contact Closures and BCD connectors
- Automatically filters sample "In-The-Vial"
- Simple to use and maintain
- Thermoelectric refrigeration option controls sample temperature from 4° C
to 40° C
- Customized software features available
709AL Specifications
Standard
- Design Type: Fixed loop, XYZ movement, Positive Displacement
- Tray Capacities: 128 (8mm) sample vials
- Injections Range: 0.06µL to 200µL
- Injections Per Vial: Up to 4 with loop size < 10µL
- Precision: Less than or equal to ±0.3% RSD, ACI test conditions
- Carryover: Less than 0.25% without rinse Less than 0.004% with rinse,
under ACI test conditions
- Keypad: 20 key, 4 line, 80 character LCD
- Valve: Valco Cheminert C3 valve with 10µL loop and high speed actuator
- Programming: 10 Methods, maximum 128 steps per method. Priority sampling
- Communications: Interfaces: RS-232-C, IEEE, I2C
- Outputs: BCD, parallel printer port, inject market, 4 user selectable NO
or NC Contact Closures with user selectable closure duration times.
- Dimensions: 33 cm H x 32 cm W x 55 cm D (13" x 13" x 21")
- Weight: 14 Kg (31 lbs.)
- Weight: 17 Kg (38 lbs.) Refrigerated
- Wetted Materials: Teflon, 316 stainless steel and nitronic 60
- Voltage: Autoselectable: 100-240 VAC ±10%, 47-63 Hz
Options
- Valves: C2, C4 or CI4W internal loop and C6 external loop Valves available
- Refrigeration: 4° C to 40° C within ±1° C accuracy using Peltier cooling
of entire tray surface
- FilterCaps: 5.0 µm, 0.45 µm and 0.2 µm polypropylene filter membranes
available
- Materials: Biocompatible materials available
- Other: Valveless models available for FIA and other direct injection
applications
Product Part
Number
709/00000/01 Standard Model 709AL with Cheminert C3 Valve with Fixed
10ul loop
709/00000/01R Refrigerated Model 709AL with C3 Valve with Fixed 10ul
loop
708/25921/00 Optional UV Cover (recommended to be purchased with
autosampler)